Ensure the weighing platform is completely empty. When the screen displays "load zero" , press the Test button. The scale will record the empty weight as its baseline.
The display will prompt for a weight (e.g., "Load 10" ). Gently place your certified 10lb (or equivalent metric) weight onto the center of the platform. Press the Test button again. avery berkel l126 calibration
For more specific configurations or if the "DIAG" menu does not appear, you can consult the official documentation on the Avery Berkel Support Page. Ensure the weighing platform is completely empty
: The display will show a default weight value (such as 160.1 ). If this does not match your intended calibration weight, use the Zero button to adjust it and confirm with the Test button . The display will prompt for a weight (e
: Use the red Zero button to cycle through units (e.g., ounces or pounds) until your preferred measurement appears, then select it with the blue Test button .